Exp 6: Dynamic Light Scattering (DLS) Hydrodynamic Sizing & Polydispersity (PDI)

Experiment 06Light Scattering

Dynamic Light Scattering (DLS): Hydrodynamic Diameter ($D_h$) & Polydispersity Index (PDI)

Stokes-Einstein Equation • Autocorrelation Function g2(τ) • CONTIN Sizing
Aim: To measure time-dependent laser intensity fluctuations from Brownian motion in a protein sample, compute the intensity autocorrelation function $g^{(2)}(tau)$, determine the translational diffusion coefficient ($D_t$), and calculate the hydrodynamic Z-average diameter ($D_h$) and Polydispersity Index (PDI) using the Stokes-Einstein equation.
๐Ÿ”ฌ Principle & Theoretical Background

Dynamic Light Scattering (DLS), or Quasi-Elastic Light Scattering (QELS), measures the rate of diffusion of macromolecules undergoing continuous, random Brownian motion in liquid suspension. When illuminated by monochromatic, coherent laser light ($lambda = 633text{ nm}$), particles scatter light that interferes constructively and destructively at a high-speed detector (Avalanche Photodiode, APD).

Because smaller particles move faster than larger particles, their scattered light intensity fluctuates on a much shorter timescale. These temporal fluctuations are quantified by the normalized second-order intensity autocorrelation function ($g^{(2)}(tau)$):
$$g^{(2)}(tau) = frac{langle I(t) cdot I(t + tau) rangle}{langle I(t) rangle^2}$$
where $tau$ is the decay delay (lag) time. Via the Siegert relation, $g^{(2)}(tau)$ relates to the electric field correlation function $g^{(1)}(tau)$:
$$g^{(2)}(tau) = 1 + beta |g^{(1)}(tau)|^2 = 1 + beta exp(-2Gamma tau)$$
where $beta$ is instrument coherence factor and $Gamma$ is the characteristic decay rate:
$$Gamma = D_t cdot q^2$$
where $D_t$ is the translational diffusion coefficient and $q$ is the magnitude of the scattering vector ($q = frac{4pi n}{lambda_0} sin(theta/2)$).

From $D_t$, the hydrodynamic diameter ($D_h$) is calculated via the fundamental Stokes-Einstein relation:
$$D_h = frac{k_B cdot T}{3pi cdot eta cdot D_t}$$
where $k_B$ is the Boltzmann constant ($1.3806 times 10^{-23}text{ J/K}$), $T$ is absolute temperature ($298.15text{ K}$), and $eta$ is solvent dynamic viscosity ($0.8872text{ cP} = 8.872 times 10^{-4}text{ Pa}cdottext{s}$ for water at $25^circtext{C}$).

Polydispersity Index (PDI): Cumulants analysis fits $ln[g^{(1)}(tau)]$ to a polynomial series: $ln[g^{(1)}(tau)] = -bar{Gamma}tau + frac{mu_2}{2}tau^2$. The dimensionless Polydispersity Index ($text{PDI} = mu_2 / bar{Gamma}^2$) measures particle size heterogeneity:

  • $text{PDI} < 0.08$: Monodisperse, high purity (suitable for protein crystallization, Cryo-EM, and NMR).
  • $0.08 le text{PDI} le 0.25$: Narrow moderately polydisperse distribution.
  • $text{PDI} > 0.30$: Broadly polydisperse, aggregate contamination.
DYNAMIC LIGHT SCATTERING (DLS): HYDRODYNAMIC DIAMETER & POLYDISPERSITY (PDI)A. Intensity Autocorrelation Function g²(τ)g²(τ) – 11.00.50.0Lag Time τ (μs, log scale)1101001000Stokes-Einstein Diffusion Equation:D_h = k_B · T / (3 · π · η · D_t)• D_t = Translational diffusion coefficient from decay rate Γ = D_t · q²• k_B = Boltzmann constant; T = Absolute temperature (298.15 K)• η = Solvent viscosity (water = 0.8872 cP at 25°C); q = scattering vectorPolydispersity Index (PDI) Standards:• PDI < 0.10: Highly Monodisperse (Crystallography / NMR Grade)• 0.10 – 0.25: Narrow Moderately Disperse | • PDI > 0.30: Aggregated / PolydisperseB. Hydrodynamic Size Distribution (CONTIN Algorithm)Scattering Intensity (%)0%10%20%30%Hydrodynamic Diameter D_h (nm, log scale)0.52.06.5 (BSA)505005000Native Monomer PeakDh = 6.48 nm ± 0.3 nmAggregate (~450 nm)Rayleigh I ∝ d⁶Sample Assessment: Monomer fraction 98.2% by volume | PDI = 0.052
๐Ÿ“‹ Reagents, Consumables & Instrumentation
  • Dynamic Light Scattering Instrument: Malvern Zetasizer Nano-ZS or DynaPro Plate Reader III ($632.8text{ nm}$ He-Ne laser, $173^circ$ Non-Invasive Backscatter NIBS optics).
  • Cuvette: Low-volume disposable polystyrene micro-cuvette ($40 mutext{L}$) or optical quartz square cell ($45 mutext{L}$, Zen2112).
  • Protein Samples: Native Bovine Serum Albumin (BSA, $M_r = 66.4text{ kDa}$) and Monoclonal IgG antibody ($M_r = 150text{ kDa}$).
  • Ultra-Clean Buffer: $20text{ mM Phosphate-Buffered Saline}$ ($text{pH } 7.4$), filtered sequentially through $0.22 mutext{m}$ and $0.02 mutext{m}$ Anotop syringe filters to eliminate particulate dust.
โš™๏ธ Step-by-Step Practical Protocol
1
Sample Microfiltration & Dust Exclusion

DLS is exceptionally sensitive to dust because Rayleigh scattering intensity scales with particle diameter to the 6th power ($I propto d^6$). Centrifuge protein stock at $16,000 times g$ for 20 minutes at $4^circtext{C}$. Filter working buffer through a $0.02 mutext{m}$ syringe filter. Pipette $50 mutext{L}$ of $1.0text{ mg/mL}$ protein sample into dust-free cuvette using clean aerosol-barrier tips.

2
Instrument Thermal Equilibration

Insert cuvette into Zetasizer sample holder. Set temperature controller to precisely $25.0^circtext{C}$ and allow thermal equilibration for 120 seconds. Enter solvent parameters: Water viscosity $eta = 0.8872text{ cP}$, refractive index $n = 1.330$, protein refractive index $1.450$.

3
Measurement Acquisition & Autocorrelation Decay

Collect 3 independent measurement runs, each consisting of 12 sub-runs of 10 seconds. The correlator accumulates raw photon count rates (target $200–400text{ kcounts/s}$). Inspect raw correlation curve: a clean exponential drop from $1.0$ to $0.0$ between $1 mutext{s}$ and $1000 mutext{s}$ confirms high data fidelity.

4
Cumulants Sizing & CONTIN Distribution Deconvolution

Analyze data via both Cumulants analysis (Z-average diameter and PDI) and the CONTIN regularized inverse Laplace transform algorithm to resolve multi-modal distributions (intensity, volume, and number distributions).

๐Ÿ“Š Representative Dynamic Light Scattering Data & Hydrodynamic Sizing
Measurement RunTranslational D_t (cm²/s)Z-Average Diameter D_h (nm)PDI (Polydispersity)% Volume (Monomer Peak)Quality Criterion
Run 16.71 × 10−⁷6.48 nm0.05199.4%Passed (Monodisperse)
Run 26.68 × 10−⁷6.51 nm0.04899.1%Passed (Monodisperse)
Run 36.73 × 10−⁷6.46 nm0.05699.2%Passed (Monodisperse)
Mean ± SD6.71 ± 0.02 × 10−⁷6.48 ± 0.03 nm0.052 ± 0.00499.2 ± 0.2%Crystallography Grade

Stokes-Einstein Calculation:
$$D_h = frac{(1.3806 times 10^{-23}text{ J/K}) times (298.15text{ K})}{3 times 3.14159 times (8.872 times 10^{-4}text{ Pa}cdottext{s}) times (6.707 times 10^{-11}text{ m}^2/text{s})} = mathbf{6.48 times 10^{-9}text{ m}} = mathbf{6.48text{ nm}}$$

Stokes Radius ($R_s$): $R_s = D_h / 2 = mathbf{3.24text{ nm}}$, in excellent agreement with size exclusion chromatography ($3.12text{ nm}$).

Structural Sample Conclusion: With a Polydispersity Index of $0.052$ ($< 0.08$) and $>99%$ monomer volume fraction, the preparation is monodisperse, aggregate-free, and suitable for structural determination via X-ray crystallography and Cryo-EM.